CVE-2026-4216

A weakness has been identified in i-SENS SmartLog App up to 2.6.8 on Android. This affects an unknown function of the component air.SmartLog.android. This manipulation causes hard-coded credentials. The attack can only be executed locally. The exploit has been made available to the public and could be used for attacks. The vendor explains: "The function referenced in the report currently exists in our deployed system. It is related to a developer mode used during the configuration process for Bluetooth pairing between the blood glucose meter and the SmartLog application. This function is intended for configuration purposes related to device integration and testing. (...) [I]n a future application update, we plan to review measures to either remove the developer mode function or restrict access to it."
Configurations

No configuration.

History

No history.

Information

Published : 2026-03-16 14:20

Updated : 2026-03-16 14:53


NVD link : CVE-2026-4216

Mitre link : CVE-2026-4216

CVE.ORG link : CVE-2026-4216


JSON object : View

Products Affected

No product.

CWE
CWE-259

Use of Hard-coded Password

CWE-798

Use of Hard-coded Credentials